NDJ-8SE
4.3-inch Touchscreen Rotational Viscometer
A touchscreen model with low-speed options and stepless adjustment for rotational viscosity measurements under a defined test procedure.

Series illustration. Appearance and supplied configuration require confirmation. Screen imagery is illustrative.Dynamic viscosity
Measight NDJ-8S is an LCD digital rotational viscometer for tests defined by a spindle, speed and sample temperature. Its listed speed settings are 0.3, 0.6, 1.5, 3, 6, 12, 30 and 60 rpm. The catalog spindle set is Nos. 1–4, with a No. 0 option for lower-viscosity work. Specify the instrument revision, container and temperature accessories when choosing a configuration.
Values apply to the stated configuration and test conditions.
| Viscosity range | 10–2,000,000 mPa·s |
|---|---|
| Preset speeds | 0.3, 0.6, 1.5, 3, 6, 12, 30, 60 |
| Spindles | 1, 2, 3, 4 |
| Display | Backlit LCD |
| Displayed values | viscosity, speed, sample temperature, torque percentage, spindle number, current maximum range |
| Spindle system | No. 1 to No. 4 included; No. 0 optional |
| Accuracy | ±2% full scale |
| Repeatability | ±1% full scale |
| Temperature measurement range | 0–99 °C |
| Sample volume | 250–400 mL |
| Operating temperature | 5–35 °C |
| Maximum relative humidity | ≤80% RH |
| Input power | AC 220 V ±10%, 50 Hz |
| Instrument dimensions | 280 × 250 × 390 mm |
| Net weight | 7 kg |
| Viscosity resolution | 1 mPa·s; 0.1 mPa·s with No. 0 spindle |
| Temperature measurement accuracy | ±0.5 °CWith the optional semiconductor temperature probe. |
| Temperature resolution | 0.1 °CWith the optional semiconductor temperature probe. |
| Spindle and speed search | Automatic search for an unknown sample |
Optional accessories include a circulating thermostatic water bath, a matching double-wall sample cup and a No. 0 low-viscosity adapter. Confirm the probe type and temperature range for the supplied instrument.
For a NDJ-8S quotation, provide the sample, expected viscosity and unit, test temperature, required method, sample volume and reporting needs. State the destination power supply and quantity. Ask for an itemised spindle, container, temperature-equipment and software configuration, together with the available model-specific documents.
Request a configuration quoteThe lowest catalog values may require the optional No. 0 spindle or matching low-viscosity adapter. Quote the spindle and container together, and confirm the usable range at the required speed; the overall instrument range is not the range of every spindle.
A temperature probe measures temperature; it does not by itself heat or cool the sample. Confirm the probe supplied with the instrument and the separate bath, cup or conditioning equipment required by the test method.
Contact our team for the current manual and configuration details.
Request documentation
4.3-inch Touchscreen Rotational Viscometer
A touchscreen model with low-speed options and stepless adjustment for rotational viscosity measurements under a defined test procedure.

LCD Digital Rotational Viscometer
A backlit digital display and nine preset speeds, including low-speed settings, with temperature measurement alongside viscosity results.