DV79B
Rotational Viscometer
Three selectable speeds and a 15 mL sample configuration for rotational viscosity measurements when sample quantity is limited.

Series illustration. Appearance and supplied configuration require confirmation. Screen imagery is illustrative.Rotational
Measight DV79A measures dynamic viscosity with a Type A container and a 70 mL sample configuration. The listed range is 1–50 mPa·s across matching spindle/speed combinations, with selectable speeds of 7.5, 75 and 750 rpm. The catalog also lists temperature measurement and RS232 output. Choose the matching cup and spindle; the overall range is not available from every combination.
Values apply to the stated configuration and test conditions.
| Rotational speeds | 7.5, 75, 750 rpm |
|---|---|
| Sample volume | 70 mL |
| Communication | RS232 |
| Torque operating range | 10–100% |
| Viscosity range | 1–50 mPa·sWith the Type A container and its matching spindles. Each spindle/speed combination covers part of this range. |
| Spindle system | A, B, C and D spindles with Type A container |
| Accuracy | ±1% full scale |
| Repeatability | ±0.5% full scale |
| Input power | 110–220 V, 50 ±1 Hz |
| Temperature probe | RTD probe |
| Drive system | Stepper motor with microprocessor control |
| Displayed values | Viscosity, speed, temperature, torque percentage and spindle number |
| Catalog temperature accuracy | 0.1 °C |
| Temperature measurement range | 0–200 °C |
Selected specifications from the current catalog are shown below. Contact us for the complete model-specific datasheet, measurement range, accessories and supplied configuration.
Request the complete datasheetUse the spindle specified for the selected container. The catalog identifies readings below 10% or above 100% torque as invalid. Record temperature and speed alongside the viscosity result.
At 750 rpm: spindle A covers 1–10 mPa·s (3,550 s⁻¹); B covers 2–20 mPa·s (1,855 s⁻¹); C covers 4–40 mPa·s (1,006 s⁻¹); D covers 5–50 mPa·s (836 s⁻¹). These shear rates apply to the listed spindle/container geometries and the 70 mL sample configuration.
Consider DV79A when the sample and method fit the Type A low-viscosity configuration. For example, the catalog lists 1–10 mPa·s for spindle A at 750 rpm with the matching 70 mL Type A setup. A different spindle or speed needs its own working-range check.
For a DV79A quotation, provide the sample, expected viscosity and unit, test temperature, required method, sample volume and reporting needs. State the destination power supply and quantity. Ask for an itemised spindle, container, temperature-equipment and software configuration, together with the available model-specific documents.
Request a configuration quoteThe RTD probe measures sample temperature. It does not by itself establish heating or cooling capability. Specify the required test temperature and confirm how the selected cup will be conditioned and held at that temperature.
Contact our team for the current manual and configuration details.
Request documentation
Rotational Viscometer
Three selectable speeds and a 15 mL sample configuration for rotational viscosity measurements when sample quantity is limited.

Small-sample Touchscreen Viscometer
A coaxial-cylinder measuring system for small sample quantities, with touchscreen operation and adjustable rotational speed.