STM-2 Touchscreen Stormer Viscometer Series — series illustrationSeries illustration. Appearance and supplied configuration require confirmation. Screen imagery is illustrative.

Stormer / Krebs

STM-2T

Stormer / Krebs Viscometer

Measight STM-2T is a touchscreen Stormer / Krebs viscometer for paint and coating tests reported in Krebs units (KU). Its listed range is 40.2–141.0 KU at 200 ±1 rpm. The instrument uses a 500 mL-capacity container, includes a temperature probe and stores 40 measurement groups. Confirm the sample filling procedure and data-connection package; container capacity is not the specified sample fill volume.

  • Krebs range: 40.2–141.0 KU
  • Rotational speed: 200 ±1 rpm
  • Container capacity: 500 mL
Request a Quote

Technical specifications

Values apply to the stated configuration and test conditions.

Krebs range40.2–141.0 KU
Rotational speed200 ±1 rpm
Container capacity500 mL
Net weight6 kg
Input power220 V, 50 Hz
Power consumption10 W
DisplayTouchscreen
Viscosity range27–5,250 cP
Load range32–1,099 g
Accuracy±1% full scale
Repeatability±1% full scale
Container dimensions93 × 85 mm
Instrument dimensions210 × 180 × 500 mm
Measurement memory40 groups
Temperature probeIncluded
Motor protectionAutomatic stop and alarm on stall or over-range
RS232 / software packageListed for STM-2T

Choose your measurement configuration.

Selected specifications from the current catalog are shown below. Contact us for the complete model-specific datasheet, measurement range, accessories and supplied configuration.

Request the complete datasheet

Repeatable sample positioning

The adjustable lifting mechanism lets the operator align the spindle immersion mark with the sample surface. The same spindle covers the listed test range. The motor stops and raises an alarm if it stalls or the measurement exceeds the range.

What to include in your STM-2T enquiry

For a STM-2T quotation, provide the sample, expected viscosity and unit, test temperature, required method, sample volume and reporting needs. State the destination power supply and quantity. Ask for an itemised spindle, container, temperature-equipment and software configuration, together with the available model-specific documents.

Request a configuration quote

Common questions

When should I consider the STM-2T?

Consider this instrument when the paint or coating procedure calls for a Stormer/Krebs measurement in KU. Match the required method, sample preparation and temperature; a KU measurement is not automatically interchangeable with a different rotational-viscometer procedure.

Does the STM-2T require 500 mL of sample?

The 500 mL figure is the container capacity, not an instruction to add 500 mL. Follow the supplied filling procedure and align the spindle immersion mark with the sample surface. Request the required fill quantity for the actual container.

Does the STM-2T temperature probe control the sample temperature?

A temperature probe measures temperature; it does not by itself heat or cool the sample. Confirm the probe supplied with the instrument and the separate bath, cup or conditioning equipment required by the test method.

Can the STM-2T export results to a computer?

An RS232/software package is listed for STM-2T. Confirm the physical interface, software licence, supported computer environment and export format in the quotation. The 40-group onboard memory alone does not specify export functions.

Documentation

Contact our team for the current manual and configuration details.

Request documentation

Related instruments

5-inch Touchscreen Rotational Viscometer Series — series illustration

NDJ-5T

Touchscreen Rotational Viscometer

A touchscreen model with stepless speed adjustment for routine rotational viscosity testing. Choose the spindle to suit your sample and method.

View details
STM-2 Touchscreen Stormer Viscometer Series — series illustration

STM-2

Stormer / Krebs Viscometer

For paint and coating consistency checks in Krebs Units, with a 40.2–141.0 KU range and a fixed-speed Stormer measurement setup.

View details